Amino Acid Separation without Derivatizing


Chart
This figure shows the extracted ion chromatogram (EIC) of a nineteen amino acid mixture under gradient conditions.

Click on Chromatogram to enlarge


Method Conditions and Sample Preparation Protocol

From our experience with amino acid separations in the ANP mode of separation, samples should be prepared as follows:
  1. Stock sample solution (if used) should be prepared in 100% water with 0.1& formic acid (or acetic acid: depends which acid is used in your mobile phase).
  2. Samples used for LC-MS should be diluted using 60% acetonitrile + 0.1%formic (or acetic acid) /40% DI water +0.1% formic (or acetic);
If the stock sample solution is not used sample should be prepared as in point 2 above.

Method Run Conditions:
Column: 2.1 x 150 mm Cogent Diamond Hydride™
Solvent A: DI water + 0.1% formic (or acetic) acid
Solvent B: Acetonitrile + 0.1% formic (or acetic acid)
Flow rate: 0.4 mL/min
Gradient conditions:
     0 min - 100 %B
     0-5 min to 75% B
     5 - 6 min to 60% B
     6 - 7 min hold 60% B
     7 - 9 min to 50%B
     9 - 10 min hold 50%B
     5 min post time at 100%B (for equlibration)
Please note Methanol is not a solvent of choice for ANP analysis, but Acetone is an excellent substitute when using LCMS.

 